FDA Adverse Event Injury Summary report: N

VIRTUOSO DR

MDR report key: 2131851 · Received June 16, 2011

Report

Report Number
6000094-2011-01178
Event Type
Injury
Date Received
June 16, 2011
Manufacturer
MEDTRONIC S.A.
Product Code
LWS
PMA / PMN Number
ASKU
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. THIS MODEL NUMBER IS NOT APPROVED FOR DISTRIBUTION IN THE UNITED STATES, HOWEVER, IT IS SAME/SIMILAR TO A DEVICE MARKETED IN THE U.S.

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. THIS MODEL NUMBER IS NOT APPROVED FOR DISTRIBUTION IN THE UNITED STATES, HOWEVER, IT IS SAME/SIMILAR TO A DEVICE MARKETED IN THE U.S. EVALUATION SUMMARY FOR (B)(4): THE ACTUAL DEVICE WAS NOT RECEIVED FOR EVALUATION. WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA. OVERSENSING, VENTRICULAR (NON-SUSTAINED TACHYCARDIA) NST LESS THAN EQUAL TO 200 MS AVERAGE V-CYCLE BETWEEN (B)(6) 2011 AND (B)(6) 2011. LEAD INTEGRITY ALERT TRIGGERED, PROGRAMMER DATA SHOWS A LEAD INTEGRITY ALERT ON (B)(6) 2011. PATIENT ALERT FOR OUT OF TOLERANCE SUBTHRESHOLD LEAD IMPEDANCE ON (B)(6) 2011.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT RECEIVED INAPPROPRIATE THERAPY DUE TO T-WAVE OVERSENSING. THE DEVICE REMAINS IN USE. IT WAS LATER NOTED THAT THE SENSITIVITY OF THE DEVICE SENSING CHANNEL WAS REPROGRAMMED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT RECEIVED INAPPROPRIATE THERAPY DUE TO T-WAVE OVERSENSING. THE DEVICE REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VIRTUOSO DR IMPLANTABLE PACEMAKER/CARDIO/DEFIB LWS MEDTRONIC S.A. D164AWG ASKU

Patients

Seq Age Sex Outcome Treatment
1 Life Threatening| O| R (B)(4) COMPETITOR IMPLANTABLE TACHY LEAD| (B)(4) IMPLANTABLE PACING LEAD| (B)(4) COMPETITOR IMPLANTABLE TACHY LEAD| (B)(4) IMPLANTABLE TACHY LEAD| (B)(4) IMPLANTABLE TACHY LEAD| (B)(4) IMPLANTABLE PACING LEAD