FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 2131850 · Received June 16, 2011

Report

Report Number
2649622-2011-10573
Event Type
Injury
Date Received
June 16, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S24
Removal / Correction Number
Z-0475-2011
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(6) THE ACTUAL DEVICE WAS NOT RECEIVED FOR EVALUATION. WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA. SENSING - OVERSENSING 1 - VENTRICULAR NST=200 MS AVERAGE V-CYCLE ON (B)(6) 2011 06:23:43. 1 - VF=160 MS ON (B)(6) 2011 AT 16:50:07. SENSING - INTERFERENCE/NOISE VENTRICULAR SHORT INTERVAL COUNT V-SIC=25.9 COUNTS AVG/DAY, IN 50.7 DAYS, BETWEEN (B)(6) 2011 17:12:20 AND (B)(6) 2011 09:36:14.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEAD EXHIBITED OVERSENSING. DURING LEAD REVISION PROCEDURE, IT WAS NOTED THAT THE PIN WAS IN THE CONNECTOR BLOCK AT AN ANGLE, AND OVERSENSING WAS NOTED WHILE MANIPULATING THE PORTION OF THE LEAD DISTAL TO THE CONNECTOR BLOCK. THE LEAD WAS REINSERTED WITHOUT ADDITIONAL OVERSENSING, AND REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6947 ASKU

Patients

Seq Age Sex Outcome Treatment
1 63 YR Hospitalization| R (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB