SPRINT QUATTRO SECURE
Report
- Report Number
- 2649622-2011-10573
- Event Type
- Injury
- Date Received
- June 16, 2011
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- LWS
- PMA / PMN Number
- P920015/S24
- Removal / Correction Number
- Z-0475-2011
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(6) THE ACTUAL DEVICE WAS NOT RECEIVED FOR EVALUATION. WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA. SENSING - OVERSENSING 1 - VENTRICULAR NST=200 MS AVERAGE V-CYCLE ON (B)(6) 2011 06:23:43. 1 - VF=160 MS ON (B)(6) 2011 AT 16:50:07. SENSING - INTERFERENCE/NOISE VENTRICULAR SHORT INTERVAL COUNT V-SIC=25.9 COUNTS AVG/DAY, IN 50.7 DAYS, BETWEEN (B)(6) 2011 17:12:20 AND (B)(6) 2011 09:36:14.
IT WAS REPORTED THAT THE LEAD EXHIBITED OVERSENSING. DURING LEAD REVISION PROCEDURE, IT WAS NOTED THAT THE PIN WAS IN THE CONNECTOR BLOCK AT AN ANGLE, AND OVERSENSING WAS NOTED WHILE MANIPULATING THE PORTION OF THE LEAD DISTAL TO THE CONNECTOR BLOCK. THE LEAD WAS REINSERTED WITHOUT ADDITIONAL OVERSENSING, AND REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPRINT QUATTRO SECURE | IMPLANTABLE TACHY LEAD | LWS | MEDTRONIC PUERTO RICO, INC. | 6947 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 63 YR | Hospitalization| R | (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB |