FDA Adverse Event Injury Summary report: N

CAPSUREFIX NOVUS

MDR report key: 2131829 · Received June 16, 2011

Report

Report Number
2649622-2011-10567
Event Type
Injury
Date Received
June 16, 2011
Date of Event
April 28, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039/S009
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE ACTUAL LEAD WAS NOT RECEIVED FOR EVALUATION. WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA. MANUAL RESET OF LEAD TREND DATA ON (B)(6) 2011. NO DATA TO REVIEW IN S2D.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT WAS IMPLANTED APPROXIMATELY (B)(6) AGO WITH A NEW PACEMAKER WITH EXISTING CHRONIC LEADS. THE PATIENT PRESENTED TO THE EMERGENCY ROOM WITH SYNCOPE AND A HEART RATE OF TEN BEATS PER MINUTE AND WAS BEING EXTERNALLY PACED. THE CHRONIC VENTRICULAR LEAD SHOWED THAT THRESHOLDS HAD RISEN. THE LEAD WAS CAPPED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 5076 ASKU

Patients

Seq Age Sex Outcome Treatment
1 73 YR Hospitalization| L| R (B)(4) IMPLANTABLE PULSE GENERATOR| (B)(4) IMPLANTABLE PULSE GENERATOR| 5076 IMPLANTABLE PACING LEAD| 5076 IMPLANTABLE PACING LEAD