FDA Adverse Event Injury Summary report: N

VIRTUOSO II DR

MDR report key: 2131814 · Received June 16, 2011

Report

Report Number
6000144-2011-03537
Event Type
Injury
Date Received
June 16, 2011
Date of Event
March 18, 2011
Manufacturer
MEDTRONIC MED REL, INC.
Product Code
LWS
PMA / PMN Number
P980016/S157
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE DEVICE WAS RETURNED AND ANALYZED AND NO ANOMALIES WERE FOUND.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A RIGHT VENTRICULAR LEAD REPLACEMENT PROCEDURE THAT THE HIGH IMPEDANCE COULD BE A HEADER ISSUE WITH THE DEVICE. THE DEVICE WAS REMOVED AND REPLACED AND THEN THERE WERE NORMAL IMPEDANCE MEASUREMENTS. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VIRTUOSO II DR IMPLANTABLE PACEMAKER/CARDIO/DEFIB LWS MEDTRONIC MED REL, INC. D274DRG ASKU

Patients

Seq Age Sex Outcome Treatment
1 60 YR Hospitalization| R 6949 IMPLANTABLE TACHY LEAD| 5076 IMPLANTABLE PACING LEAD