ASKU
Report
- Report Number
- 2183613-2011-00320
- Event Type
- Malfunction
- Date Received
- June 16, 2011
- Manufacturer
- MEDTRONIC MILACA INC.
- Product Code
- DTE
- PMA / PMN Number
- P820003/S72
- Removal / Correction Number
- Z-0430-2011
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4): ANALYSIS COULD NOT CONFIRM THE CUSTOMER COMMENT. ANALYSIS DID FIND THAT THE UPPER AND LOWER CASES WERE BROKEN, THE BATTERY RELEASE WAS CONTAMINATED, ONE CONTROL KNOB WAS BROKEN, THE RING COVER WAS CONTAMINATED, TWO SIDE BAIL COVERS WERE BROKEN, THE BATTERY CONTACTS WERE OUT OF SPECIFICATION, THE RING WAS BENT, ONE SIDE BAIL WAS MISSING AND THE ENCODER FLEX AND BATTERY FLEX WERE OUT OF SPECIFICATION.
IT WAS REPORTED THAT A SELF-TEST ERROR OCCURRED. NO PATIENT INVOLVEMENT OR COMPLICATIONS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ASKU | EXTERNAL PULSE GENERATOR | DTE | MEDTRONIC MILACA INC. | 5388 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |