FDA Adverse Event Injury Summary report: N

CAPSURE SP

MDR report key: 2131670 · Received June 16, 2011

Report

Report Number
2649622-2011-10671
Event Type
Injury
Date Received
June 16, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P830061/S12
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR LEAD HAD LOW IMPEDANCES AND HIGH THRESHOLDS. THE LEAD WAS CAPPED AND REPLACED. IT WAS ALSO REPORTED THAT, PRIOR TO BEING CAPPED, THERE WAS A VENTRICULAR HIGH RATE EPISODE SHOWING NOISE ON THE LEAD. THE LEAD ALSO HAD HIGH UNIPOLAR IMPEDANCE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSURE SP IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 4024 ASKU

Patients

Seq Age Sex Outcome Treatment
1 67 YR Hospitalization| R 4068 IMPLANTABLE PACING LEAD| E2DR01 IMPLANTABLE PULSE GENERATOR