FDA Adverse Event
Malfunction
Summary report: N
CARELINK PROGRAM W/TELEMETRY C
MDR report key: 2131652
·
Received June 16, 2011
Report
- Report Number
- 2182208-2011-01003
- Event Type
- Malfunction
- Date Received
- June 16, 2011
- Manufacturer
- MEDTRONIC, INC.
- Product Code
- DXY
- PMA / PMN Number
- P890003/S080
- Removal / Correction Number
- ASKU
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) ANALYSIS COULD NOT CONFIRM THE REPORTED EVENT. INCOMING DEVICE INSPECTION FOUND THAT THE APPROPRIATE BOX WAS NOT CHECKED FOR WIRELESS TELEMETRY. (B)(4) THE STYLUS WAS FOUND TO BE DAMAGED, AND THE KEYBOARD WAS BROKEN.
Description of Event or Problem · 1
IT WAS REPORTED THAT WIRELESS TELEMETRY COULD NOT BE ESTABLISHED. NO PATIENT INVOLVEMENT OR COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CARELINK PROGRAM W/TELEMETRY C | PROGRAMMER | DXY | MEDTRONIC, INC. | 2090W | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other | 2290 PACING SYSTEM ANALYZER |