FDA Adverse Event Malfunction Summary report: N

CARELINK PROGRAM W/TELEMETRY C

MDR report key: 2131652 · Received June 16, 2011

Report

Report Number
2182208-2011-01003
Event Type
Malfunction
Date Received
June 16, 2011
Manufacturer
MEDTRONIC, INC.
Product Code
DXY
PMA / PMN Number
P890003/S080
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) ANALYSIS COULD NOT CONFIRM THE REPORTED EVENT. INCOMING DEVICE INSPECTION FOUND THAT THE APPROPRIATE BOX WAS NOT CHECKED FOR WIRELESS TELEMETRY. (B)(4) THE STYLUS WAS FOUND TO BE DAMAGED, AND THE KEYBOARD WAS BROKEN.

Description of Event or Problem · 1

IT WAS REPORTED THAT WIRELESS TELEMETRY COULD NOT BE ESTABLISHED. NO PATIENT INVOLVEMENT OR COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CARELINK PROGRAM W/TELEMETRY C PROGRAMMER DXY MEDTRONIC, INC. 2090W ASKU

Patients

Seq Age Sex Outcome Treatment
1 Other 2290 PACING SYSTEM ANALYZER