FDA Adverse Event Malfunction Summary report: N

CARELINK PROGRAM W/TELEMETRY C

MDR report key: 2131646 · Received June 16, 2011

Report

Report Number
2182208-2011-01000
Event Type
Malfunction
Date Received
June 16, 2011
Manufacturer
MEDTRONIC, INC.
Product Code
DXY
PMA / PMN Number
P890003/S080
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BASED SOLELY ON DEVICE RETURN AND ANALYSIS. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) ANALYSIS CONFIRMED THE REPORTED EVENT. NEITHER USB OR COM PORTS WORKED. PC BOARD TESTED OUT OF SPECIFICATION. THE STYLUS FOUND DAMAGED.

Description of Event or Problem · 1

IT WAS REPORTED THAT NEITHER THE USB OR "COM" PORT WOULD WORK TO PRINT TO AN EXTERNAL PRINTER. THE PROGRAMMER WAS RETURNED TO THE MANUFACTURER, ANALYZED AND TESTED OUT OF SPECIFICATION. NO PATIENT INVOLVEMENT OR COMPLICATIONS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CARELINK PROGRAM W/TELEMETRY C PROGRAMMER DXY MEDTRONIC, INC. 2090W ASKU

Patients

Seq Age Sex Outcome Treatment
1 Other 2067 PROGRAMMER RF HEAD