FDA Adverse Event
Malfunction
Summary report: N
CARELINK
MDR report key: 2131642
·
Received June 16, 2011
Report
- Report Number
- 2182208-2011-00998
- Event Type
- Malfunction
- Date Received
- June 16, 2011
- Manufacturer
- MEDTRONIC, INC.
- Product Code
- KRG
- PMA / PMN Number
- P890003/S065
- Removal / Correction Number
- ASKU
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
Additional Manufacturer Narrative · 1
THIS REPORT IS BASED SOLELY ON DEVICE RETURN AND ANALYSIS. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: ANALYSIS CONFIRMED THE REPORTED EVENT. (B)(4) RF HEAD UPLINK WAS OUT OF SPECIFICATION. (B)(4) STYLUS AND PRINTER DRAWER FOUND BROKEN. SYSTEM FAN IS NOISY.
Description of Event or Problem · 1
IT WAS REPORTED TELEMETRY IS INCONSISTENT, DESPITE CHANGING THE RF HEAD. THE PROGRAMMER AND RF HEAD WERE RETURNED TO THE MANUFACTURER, ANALYZED AND TESTED OUT OF SPECIFICATION. NO PATIENT INVOLVEMENT OR COMPLICATIONS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CARELINK | PROGRAMMER | KRG | MEDTRONIC, INC. | 2090 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |