FDA Adverse Event Malfunction Summary report: N

SPRINT QUATTRO SECURE S

MDR report key: 2131633 · Received June 16, 2011

Report

Report Number
2649622-2011-10529
Event Type
Malfunction
Date Received
June 16, 2011
Date of Event
April 29, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S039
Removal / Correction Number
Z-0474-2011
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. ANALYSIS OF THE LEAD IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE.

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY: (B)(4), THE FULL LEAD WAS RETURNED AND ANALYZED. NO ANOMALIES WERE FOUND. IT WAS NOTED THE INNER TUBING WAS KINKED/BUCKLED, THERE WAS BLOOD IN/ON HELIX/LOBE MECHANISM (SLEEVE HEAD) AND HELIX/LOBE MECHANISM, AND THE LEAD APPEARED TO BE DAMAGED AT IMPLANT. THE ANALYST ALSO NOTED DRIED BLOOD AND TISSUE ON THE HELIX.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN IMPLANT, THE LEAD WAS ATTEMPTED DUE TO HIGH IMPEDANCE AND THRESHOLD VALUES. THE LEAD WAS REMOVED AND REPLACED WITH A DIFFERENT LEAD. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE S IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6935 ASKU

Patients

Seq Age Sex Outcome Treatment
1 83 YR Other