FDA Adverse Event
Malfunction
Summary report: N
CARELINK PROGRAM W/TELEMETRY C
MDR report key: 2131609
·
Received June 16, 2011
Report
- Report Number
- 2182208-2011-00993
- Event Type
- Malfunction
- Date Received
- June 16, 2011
- Manufacturer
- MEDTRONIC, INC.
- Product Code
- DXY
- PMA / PMN Number
- P890003/S080
- Removal / Correction Number
- ASKU
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
Additional Manufacturer Narrative · 1
THIS REPORT IS BASED SOLELY ON DEVICE RETURN AND ANALYSIS. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) ANALYSIS CONFIRMED THE REPORTED EVENT. THE PROGRAMMER POWERED UP WITH A SYSTEM ERROR. THE STYLUS AND THE POWER CORD DOOR WERE ALSO FOUND TO BE BROKEN, AND THE SYSTEM FAN WAS MAKING LOUD NOISE.
Description of Event or Problem · 1
IT WAS REPORTED THE PROGRAMMER HAD A BLACK SCREEN AND SERIOUS PROBLEM ERROR. THE PROGRAMMER WAS RETURNED TO THE MANUFACTURER, ANALYZED, AND TESTED OUT OF SPECIFICATION. NO PATIENT INVOLVEMENT OR COMPLICATIONS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CARELINK PROGRAM W/TELEMETRY C | PROGRAMMER | DXY | MEDTRONIC, INC. | 2090W | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other | 2290 PACING SYSTEM ANALYZER |