MENTOR MEMORYSHAPE BREAST IMPLANT
Report
- Report Number
- 1645337-2025-01167
- Event Type
- Injury
- Date Received
- February 5, 2025
- Date of Event
- January 21, 2025
- Manufacturer
- MENTOR TEXAS
- Product Code
- FTR
- UDI-DI
- 00081317005193
- PMA / PMN Number
- P060028
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OK, US
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
ON 20-MAR-2025, MENTOR RECEIVED ADDITIONAL INFORMATION ABOUT THE EVENT. IT WAS REPORTED THAT THE LOT NUMBERS AND SERIAL NUMBERS FOR THE PATIENT¿S BREAST IMPLANTS WERE INADVERTENTLY MIXED UP. THE CORRECT LOT NUMBER FOR THE PATIENT¿S RUPTURED LEFT BREAST IMPLANT IS 7003845 AND THE CORRECT SERIAL NUMBER IS (B)(6). THE DEVICE IS A MENTOR MEMORYSHAPE BREAST IMPLANT 345CC GEL BREAST PROSTHESIS AND CATALOG NUMBER AND UDI HAVE BEEN UPDATED. ON 21-MAR-2025, MENTOR COMPLETED THE INVESTIGATION ON THE DEVICE FROM LOT #7003845. DEVICE EVALUATION SUMMARY: ACCORDING TO THE INFORMATION RECEIVED, THE PATIENT EXPERIENCED A RUPTURE IN THE BREAST IMPLANT. THE PRODUCT WAS RETURNED TO MENTOR FOR EVALUATION. MENTOR THEN CONDUCTED A VISUAL INSPECTION AND MICROSCOPIC EXAMINATION OF THE RETURNED DEVICE. VISUAL ANALYSIS OF THE RETURNED DEVICE DETERMINED THAT THE BREAST IMPLANT HAD A TEAR ON THE ANTERIOR VIEW, MEASURING APPROXIMATELY 6.5 CM. MICROSCOPIC EXAMINATION WAS PERFORMED ON THE EDGES OF THE RUPTURE, AND PARALLEL STRIATIONS WERE FOUND IN AN AREA OF THE TEAR, MEASURING APPROXIMATELY 3.0 CM. PARALLEL STRIATIONS ARE CONSISTENT WITH MARKINGS MADE BY A SHARP OBJECT PERFORATING THE IMPLANT SHELL. THE CAUSE OF THE RUPTURE IN THE REMAINING AREA OF THE TEAR COULD NOT BE IDENTIFIED. MENTOR ALSO PERFORMED A MANUFACTURING RECORD EVALUATION RELATED TO THE REPORTED COMPLAINT FOR THE FINISHED DEVICE LOT NUMBER. NO MANUFACTURING NON-CONFORMANCES WERE IDENTIFIED AS PART OF THIS EVALUATION. THE CONTRALATERAL DEVICE WAS ALSO RECEIVED (LOT NUMBER 7549786). THE PATIENT DID NOT REPORT ANY ISSUES WITH THIS DEVICE. THEREFORE, NO FURTHER ANALYSIS OF THE CONTRALATERAL DEVICE IS REQUIRED. AS PART OF MENTOR'S QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED ACCORDING TO APPROVED SPECIFICATIONS. BASED ON THE RESULTS OF THIS INVESTIGATION, IT WAS DETERMINED THAT THE PRODUCT MET THE MANUFACTURING RELEASE CRITERIA: NO CORRECTIVE AND PREVENTIVE ACTION (CAPA) IS REQUIRED NOW. ADDITIONAL COMPLAINT INFORMATION MONITORING FOR POTENTIAL SAFETY SIGNALS IS CONDUCTED THROUGH COMPLAINT TRENDING AS PART OF POST-MARKET SURVEILLANCE. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).
ON 18-FEB-2025, THE MENTOR FAILURE ANALYSIS LAB RECEIVED THE DEVICE FOR EVALUATION. ON 25-FEB-2025, MENTOR COMPLETED THE INVESTIGATION ON THE SUSPECT MEDICAL DEVICE. DEVICE EVALUATION SUMMARY: ACCORDING TO THE INFORMATION RECEIVED, THE PATIENT EXPERIENCED A RUPTURE IN THE BREAST IMPLANT. THE PRODUCT WAS RETURNED TO MENTOR FOR EVALUATION. MENTOR THEN CONDUCTED A VISUAL INSPECTION AND MICROSCOPIC EXAMINATION OF THE RETURNED DEVICE. VISUAL ANALYSIS OF THE RETURNED DEVICE DETERMINED THAT THE BREAST IMPLANT WAS FOUND TO BE RUPTURED AND IT WAS RECEIVED IN THREE (3) PARTS. MICROSCOPIC EXAMINATION WAS PERFORMED ON THE EDGES OF THE RUPTURE, AND PARALLEL STRIATIONS WERE FOUND IN AN AREA OF THE TEAR ON THE ANTERIOR ASPECT, MEASURING APPROXIMATELY LESS THAN 0.1 CM. PARALLEL STRIATIONS ARE CONSISTENT WITH MARKINGS MADE BY A SHARP OBJECT PERFORATING THE IMPLANT SHELL. THE CAUSE OF THE RUPTURE IN THE REMAINING AREA OF THE TEAR COULD NOT BE IDENTIFIED. MENTOR ALSO PERFORMED A MANUFACTURING RECORD EVALUATION RELATED TO THE REPORTED COMPLAINT FOR THE FINISHED DEVICE LOT NUMBER. NO MANUFACTURING NON-CONFORMANCES WERE IDENTIFIED AS PART OF THIS EVALUATION. THE CONTRALATERAL DEVICE WAS ALSO RECEIVED (LOT NUMBER 7003845). THE PATIENT DID NOT REPORT ANY ISSUES WITH THIS DEVICE. THEREFORE, NO FURTHER ANALYSIS OF THE CONTRALATERAL DEVICE IS REQUIRED. AS PART OF MENTOR'S QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED ACCORDING TO APPROVED SPECIFICATIONS. BASED ON THE RESULTS OF THIS INVESTIGATION, IT WAS DETERMINED THAT THE PRODUCT MET THE MANUFACTURING RELEASE CRITERIA: NO CORRECTIVE AND PREVENTIVE ACTION (CAPA) IS REQUIRED NOW. ADDITIONAL COMPLAINT INFORMATION MONITORING FOR POTENTIAL SAFETY SIGNALS IS CONDUCTED THROUGH COMPLAINT TRENDING AS PART OF POST-MARKET SURVEILLANCE. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).
SINCE THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS CAN BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, THE INVESTIGATION WILL BE REOPENED AND CONDUCTED AS APPROPRIATE. A MANUFACTURING RECORD EVALUATION (MRE) WAS PERFORMED, AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE MRE VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. REASON FOR DEVICE EXPLANT AND/OR REOPERATION: IMPLANT EXCHANGE. MENTOR IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH MENTOR HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, MENTOR, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, MENTOR, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF CERTAIN INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).
IT WAS REPORTED THAT A PATIENT UNDERWENT A PRIMARY BREAST AUGMENTATION PROCEDURE WITH A MENTOR MEMORYSHAPE BREAST IMPLANT 390CC GEL BREAST PROSTHESIS THAT RUPTURED POST IMPLANTATION. RUPTURE OF THE PATIENT¿S LEFT BREAST PROSTHESIS WAS OBSERVED DURING AN IMPLANT EXCHANGE SURGERY ON (B)(6) 2025. THE PATIENT HAD BOTH BREAST PROSTHESES EXPLANTED AND THEY WERE REPLACED WITH ALLERGAN GEL BREAST PROSTHESES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 903853 | MENTOR MEMORYSHAPE BREAST IMPLANT | PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED | FTR | MENTOR TEXAS | 7003845 | 00081317005193 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 40 YR | Female | Required Intervention |