FDA Adverse Event
Injury
Summary report: N
CAPSUREFIX NOVUS
MDR report key: 2131563
·
Received June 16, 2011
Report
- Report Number
- 2649622-2011-10515
- Event Type
- Injury
- Date Received
- June 16, 2011
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- DTB
- PMA / PMN Number
- P930039/S009
- Removal / Correction Number
- ASKU
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SP
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS.
Description of Event or Problem · 1
IT WAS REPORTED THAT ATRIAL HIGH THRESHOLDS WERE NOTED. UPON FURTHER INVESTIGATION, THE ATRIAL LEAD AS FOUND TO HAVE DISLODGED. DURING REPOSITION, IT WAS FOUND THAT THE HELIX OF THE LEAD DID NOT WORK. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CAPSUREFIX NOVUS | IMPLANTABLE PACING LEAD | DTB | MEDTRONIC PUERTO RICO, INC. | 5076 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 28 YR | Hospitalization| R | 6947 IMPLANTABLE TACHY LEAD| (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB |