PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM
Report
- Report Number
- 2024168-2011-04254
- Event Type
- Injury
- Date Received
- June 16, 2011
- Date of Event
- April 29, 2011
- Report Date
- May 25, 2011
- Manufacturer
- AV-TEMECULA-CT
- Product Code
- MGB
- PMA / PMN Number
- P960043
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BE
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. THE DEVICE WAS NOT RETURNED FOR ANALYSIS. WITHOUT DEVICE EVALUATION, A CAUSE FOR THE UNUSUAL SOUND AND CUFF MISS COULD NOT BE DETERMINED. HOWEVER, A CUFF-MISS CAN BE INFLUENCED BY MANY FACTORS INCLUDING, BUT NOT LIMITED TO, MANUFACTURING, FAILURE TO POSITION AND MAINTAIN THE DEVICE AT A 45 DEGREE ANGLE THROUGHOUT DEPLOYMENT, ROTATING THE DEVICE AT ANY POINT DURING PLUNGER/NEEDLE DEPLOYMENT, DEPLOYMENT IN A CHALLENGING ANATOMY, AGGRESSIVELY DEPLOYING OR REMOVING THE PLUNGER AND/OR INADEQUATE POSITIONING OF THE FOOT AGAINST THE ARTERIAL WALL. WHETHER THESE CONDITIONS PLAYED A ROLE IN THE EXPERIENCED EVENT CANNOT BE VERIFIED. A REVIEW OF THE FINISHED PRODUCT LOT HISTORY RECORD DID NOT REVEAL ANY MANUFACTURING RELATED NONCONFORMING MATERIAL RECORDS ASSOCIATED WITH THIS LOT. BASED ON THE INFORMATION AVAILABLE, THE INSPECTION CRITERIA AND THE REVIEW OF THE LOT HISTORY RECORD, THERE DOES NOT APPEAR TO BE ANY INDICATIONS OF A PRODUCT QUALITY DEFICIENCY. TO ENSURE THIS TYPE OF EXPERIENCE IS NOT A RESULT OF A POTENTIAL PRODUCT RELATED DEFICIENCY, IN-PROCESS TESTING OF DEVICES ARE PERFORMED TO ASSURE THE TRAJECTORY OF NEEDLES IN ORDER TO PREVENT THIS MODE OF FAILURE. IN ADDITION, A SAMPLING OF FINISHED DEVICES IS TESTED TO VERIFY THE FUNCTIONALITY OF THE DEVICE.
IT WAS REPORTED THAT A PHYSICIAN, TRAINED IN THE USE OF THE PERCLOSE PROGLIDE, ATTEMPTED ARTERIOTOMY CLOSURE OF THE COMMON FEMORAL ARTERY AFTER A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY OF THE ILLIAC ARTERY. REPORTEDLY, DURING DEPLOYMENT OF THE NEEDLES, AN UNUSUAL SOUND WAS HEARD WHEN THE NEEDLE TIP CLICKED IN THE FOOT. A CUFF MISS OCCURRED. ANOTHER PROGLIDE WAS USED TO ACHIEVE HEMOSTASIS. THERE WAS NO ADVERSE PATIENT SEQUELA. THOUGH REQUESTED, NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM | SUTURE MEDIATED CLOSURE | MGB | AV-TEMECULA-CT | 020196H |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention | GUIDE WIRE: TERUMO .035SHEATH: CORDIS 6F |