FDA Adverse Event Injury Summary report: N

PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM

MDR report key: 2131467 · Received June 16, 2011

Report

Report Number
2024168-2011-04254
Event Type
Injury
Date Received
June 16, 2011
Date of Event
April 29, 2011
Report Date
May 25, 2011
Manufacturer
AV-TEMECULA-CT
Product Code
MGB
PMA / PMN Number
P960043
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BE
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. THE DEVICE WAS NOT RETURNED FOR ANALYSIS. WITHOUT DEVICE EVALUATION, A CAUSE FOR THE UNUSUAL SOUND AND CUFF MISS COULD NOT BE DETERMINED. HOWEVER, A CUFF-MISS CAN BE INFLUENCED BY MANY FACTORS INCLUDING, BUT NOT LIMITED TO, MANUFACTURING, FAILURE TO POSITION AND MAINTAIN THE DEVICE AT A 45 DEGREE ANGLE THROUGHOUT DEPLOYMENT, ROTATING THE DEVICE AT ANY POINT DURING PLUNGER/NEEDLE DEPLOYMENT, DEPLOYMENT IN A CHALLENGING ANATOMY, AGGRESSIVELY DEPLOYING OR REMOVING THE PLUNGER AND/OR INADEQUATE POSITIONING OF THE FOOT AGAINST THE ARTERIAL WALL. WHETHER THESE CONDITIONS PLAYED A ROLE IN THE EXPERIENCED EVENT CANNOT BE VERIFIED. A REVIEW OF THE FINISHED PRODUCT LOT HISTORY RECORD DID NOT REVEAL ANY MANUFACTURING RELATED NONCONFORMING MATERIAL RECORDS ASSOCIATED WITH THIS LOT. BASED ON THE INFORMATION AVAILABLE, THE INSPECTION CRITERIA AND THE REVIEW OF THE LOT HISTORY RECORD, THERE DOES NOT APPEAR TO BE ANY INDICATIONS OF A PRODUCT QUALITY DEFICIENCY. TO ENSURE THIS TYPE OF EXPERIENCE IS NOT A RESULT OF A POTENTIAL PRODUCT RELATED DEFICIENCY, IN-PROCESS TESTING OF DEVICES ARE PERFORMED TO ASSURE THE TRAJECTORY OF NEEDLES IN ORDER TO PREVENT THIS MODE OF FAILURE. IN ADDITION, A SAMPLING OF FINISHED DEVICES IS TESTED TO VERIFY THE FUNCTIONALITY OF THE DEVICE.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PHYSICIAN, TRAINED IN THE USE OF THE PERCLOSE PROGLIDE, ATTEMPTED ARTERIOTOMY CLOSURE OF THE COMMON FEMORAL ARTERY AFTER A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY OF THE ILLIAC ARTERY. REPORTEDLY, DURING DEPLOYMENT OF THE NEEDLES, AN UNUSUAL SOUND WAS HEARD WHEN THE NEEDLE TIP CLICKED IN THE FOOT. A CUFF MISS OCCURRED. ANOTHER PROGLIDE WAS USED TO ACHIEVE HEMOSTASIS. THERE WAS NO ADVERSE PATIENT SEQUELA. THOUGH REQUESTED, NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM SUTURE MEDIATED CLOSURE MGB AV-TEMECULA-CT 020196H

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention GUIDE WIRE: TERUMO .035SHEATH: CORDIS 6F