FDA Adverse Event
Malfunction
Summary report: N
ASKU
MDR report key: 2131463
·
Received June 16, 2011
Report
- Report Number
- 2183613-2011-00284
- Event Type
- Malfunction
- Date Received
- June 16, 2011
- Manufacturer
- MEDTRONIC MILACA INC.
- Product Code
- DTE
- PMA / PMN Number
- P820003/S72
- Removal / Correction Number
- Z-0430-2011
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
THIS REPORT IS BASED SOLELY ON DEVICE RETURN AND ANALYSIS. NO INFORMATION TO SUGGEST A DEVICE-RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4): THE DEVICE WAS RETURNED, ANALYZED AND ANALYSIS FOUND THAT THE BATTERY RELEASE, RELEASE SPRING, BATTERY DRAWER AND BATTERY FLEX WERE CONTAMINATED. IT WAS ALSO NOTED THAT THE UPPER AND LOWER CASES WERE BROKEN.
Description of Event or Problem · 1
THE DEVICE WAS RETURNED TO THE MANUFACTURER, ANALYZED, AND TESTED OUT OF SPECIFICATION. NO PATIENT INVOLVEMENT OR COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ASKU | EXTERNAL PULSE GENERATOR | DTE | MEDTRONIC MILACA INC. | 5388 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |