ASKU
Report
- Report Number
- 2183613-2011-00283
- Event Type
- Injury
- Date Received
- June 16, 2011
- Manufacturer
- MEDTRONIC MILACA INC.
- Product Code
- DTE
- PMA / PMN Number
- P820003/S72
- Removal / Correction Number
- ASKU
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- OTHER
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4): ANALYSIS COULD NOT CONFIRM THE CUSTOMER COMMENT, HOWEVER ANALYSIS DID FIND THAT THE BATTERY DRAWER WAS BROKEN, THE BATTERY CONTACTS WERE COMPRESSED, AND THE BATTERY RELEASE AND LEAD FLEX COVER WERE CONTAMINATED. IT WAS ALSO NOTED THAT THE UPPER AND LOWER CASES WERE BROKEN, THE RING COVER AND TWO SIDE BAIL COVERS WERE BROKEN, THE RING WAS BENT AND THE KEYBOARD PAD WAS COSMETICALLY SCRATCHED.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE.
IT WAS REPORTED THE EPG (EXTERNAL PULSE GENERATOR) QUIT PACING SEVERAL TIMES ON A PACING DEPENDENT PATIENT. THE EPG WAS REPLACED. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ASKU | EXTERNAL PULSE GENERATOR | DTE | MEDTRONIC MILACA INC. | 5388 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Life Threatening| R |