FDA Adverse Event Malfunction Summary report: N

ASKU

MDR report key: 2131426 · Received June 16, 2011

Report

Report Number
2183613-2011-00273
Event Type
Malfunction
Date Received
June 16, 2011
Manufacturer
MEDTRONIC MILACA INC.
Product Code
DTE
PMA / PMN Number
P820003/S72
Removal / Correction Number
Z-0430-2011
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BASED SOLELY ON DEVICE RETURN AND ANALYSIS. NO INFORMATION TO SUGGEST A DEVICE-RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4): THE DEVICE WAS RETURNED, ANALYZED AND ANALYSIS FOUND THAT THE KEYBOARD WAS CONTAMINATED. IT WAS ALSO NOTED THAT ONE SIDE BAIL COVER WAS BROKEN.

Description of Event or Problem · 1

THE DEVICE WAS RETURNED TO THE MANUFACTURER DUE TO A FIELD ADVISORY. IT WAS ANALYZED AND TESTED OUT OF SPECIFICATION. NO PATIENT INVOLVEMENT OR COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ASKU EXTERNAL PULSE GENERATOR DTE MEDTRONIC MILACA INC. 5388 ASKU

Patients

Seq Age Sex Outcome Treatment
1 Other