FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 2131404 · Received June 16, 2011

Report

Report Number
2649622-2011-10491
Event Type
Injury
Date Received
June 16, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S24
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY FOR (B)(4): THE PARTIAL LEAD WAS RETURNED IN SEGMENTS, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THE DEFIBRILLATION COIL WAS DISTORTED, THERE WAS BLOOD/BODY FLUID ON SEVERAL CONDUCTORS (NOT OBSTRUCTED), THE OUTER TUBING OVERLAY MELTED, THE OUTER TUBING OVERLAY HAD COSMETIC ENVIRONMENTAL STRESS CRACKING, THERE WAS A WHITE SUBSTANCE ON THE EXPOSED DEFIBRILLATION COIL, THE OUTER INSULATION HAD A COSMETIC DEPRESSION, THERE WAS BLOOD IN/ON THE HELIX MECHANISM (SLEEVE HEAD), THERE WAS BLOOD IN/ON THE HELIX MECHANISM AND THERE WAS APPARENT EXPLANT DAMAGE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR LEAD WAS T WAVE OVERSENSING AND VENTRICULAR PACING WAS INHIBITED. THE PATIENT'S LEFT SUBCLAVIAN IS OCCLUDED SO THE LEAD WAS REMOVED AND A NEW LEAD IMPLANTED ON THE RIGHT SIDE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6947 ASKU

Patients

Seq Age Sex Outcome Treatment
1 73 YR Hospitalization| R 5076 IMPLANTABLE PACING LEAD| 4194 IMPLANTABLE PACING LEAD| (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB