FDA Adverse Event Malfunction Summary report: N

ASKU

MDR report key: 2131384 · Received June 16, 2011

Report

Report Number
2183613-2011-00264
Event Type
Malfunction
Date Received
June 16, 2011
Manufacturer
MEDTRONIC MILACA INC.
Product Code
DTE
PMA / PMN Number
P820003/S72
Removal / Correction Number
Z-0430-2011
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BASED SOLELY ON DEVICE RETURN AND ANALYSIS. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) BATTERY CONTACTS FOUND COMPRESSED. LCD DISPLAY OUT OF SPECIFICATION (SEGMENT MISSING), RING BENT, AND UPPER AND LOWER CASES BROKEN.

Description of Event or Problem · 1

THE EXTERNAL PULSE GENERATOR WAS RETURNED TO THE MANUFACTURER FOR REPAIR, ANALYZED AND TESTED OUT OF SPECIFICATION. NO PATIENT INVOLVEMENT OR COMPLICATIONS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ASKU EXTERNAL PULSE GENERATOR DTE MEDTRONIC MILACA INC. 5388 ASKU

Patients

Seq Age Sex Outcome Treatment
1 Other