FDA Adverse Event Injury Summary report: N

CAPSUREFIX NOVUS MRI

MDR report key: 2131383 · Received June 16, 2011

Report

Report Number
2649622-2011-10484
Event Type
Injury
Date Received
June 16, 2011
Date of Event
April 8, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
NVN
PMA / PMN Number
P090013
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE PROXIMAL SEGMENT OF THE LEAD WAS RETURNED AND ANALYSIS FOUND NO ANOMALIES. IT WAS NOTED THAT ALL CONDUCTORS HAD BLOOD/BODY FLUID (NOT OBSTRUCTED), THE INNER INSULATION WAS KINKED BUCKLED AND THERE WAS APPARENT EXPLANT DAMAGE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEAD HAD NO CAPTURE, DISLODGED, AND HAD PERFORATED. THE LEAD WAS EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS MRI IMPLANTABLE PACING LEAD NVN MEDTRONIC PUERTO RICO, INC. 5086MRI ASKU

Patients

Seq Age Sex Outcome Treatment
1 43 YR Hospitalization| L| R (B)(4) IMPLANTABLE PACING LEAD| (B)(4) IMPLANTABLE PULSE GENERATOR