FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 2131355 · Received June 16, 2011

Report

Report Number
2649622-2011-10475
Event Type
Injury
Date Received
June 16, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S24
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CT, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY FOR (B)(4) THE ACTUAL DEVICE WAS NOT RECEIVED FOR EVALUATION. WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA. VARYING RESISTANCE/IMPEDANCE, WEEKLY PACING IMPEDANCE TREND DATA SHOWS VARYING IMPEDANCE FOR MIN AND MAX RV PACE EQUAL TO 664 TO 1920 OHMS RANGE BETWEEN (B)(6) 2010 AND (B)(6) 2011.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD IMPEDANCE WAS HIGH AND THERE WAS A POSSIBLE FRACTURE. THE LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6947 ASKU

Patients

Seq Age Sex Outcome Treatment
1 60 YR Hospitalization| R 4574 IMPLANTABLE PACING LEAD| 4194 IMPLANTABLE PACING LEAD| C154DWK IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 4574 IMPLANTABLE PACING LEAD| 4194 IMPLANTABLE PACING LEAD| C154DWK IMPLANTABLE PACEMAKER/CARDIO/DEFIB