FDA Adverse Event Injury Summary report: N

ATTAIN OTW

MDR report key: 2131243 · Received June 16, 2011

Report

Report Number
2649622-2011-10449
Event Type
Injury
Date Received
June 16, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P010015/S003
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY FOR (B)(4): THE FULL LEAD WAS RETURNED, ANALYZED AND THE DISTAL CONDUCTOR WAS FRACTURED. IT WAS NOTED THAT THERE WAS BLOOD/BODY FLUID ON THE DISTAL CONDUCTOR (NOT OBSTRUCTED), THE OUTER INSULATION WAS BREACHED CUT AND THE OUTER INSULATION HAD A COSMETIC DEPRESSION.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEFT VENTRICULAR (LV) LEAD HAD BEEN "PROGRAMMED OFF" PREVIOUSLY FOR HIGH IMPEDANCE AND NO CAPTURE. THE PHYSICIAN HAD NOTED THAT "AT SOME POINT THE PATIENT [HAD] INSERTED A KNIFE INTO [THE] DEVICE POCKET" WHICH COULD HAVE "POTENTIALLY DAMAGED THE LV" LEAD. IT WAS FURTHER REPORTED THAT THE LV LEAD EXPERIENCED AN APPARENT FRACTURE AND HIGH THRESHOLDS. THE LEAD WAS REMOVED AND REPLACED WITH A NEW LEAD. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEFT VENTRICULAR (LV) LEAD HAD BEEN "PROGRAMMED OFF" PREVIOUSLY FOR HIGH IMPEDANCE AND NO CAPTURE. THE PHYSICIAN HAD NOTED THAT "AT SOME POINT THE PATIENT [HAD] INSERTED A KNIFE INTO [THE] DEVICE POCKET" WHICH COULD HAVE "POTENTIALLY DAMAGED THE LV" LEAD. IT WAS FURTHER REPORTED THAT THE LV LEAD EXPERIENCED AN APPARENT FRACTURE AND HIGH THRESHOLDS. THE LEAD WAS REMOVED AND REPLACED WITH A NEW LEAD. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ATTAIN OTW IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 4193 ASKU

Patients

Seq Age Sex Outcome Treatment
1 61 YR Hospitalization| R 5076 IMPLANTABLE PACING LEAD| (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 6949 IMPLANTABLE TACHY LEAD| (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 5076| 5076 IMPLANTABLE PACING LEAD| 6949 IMPLANTABLE TACHY LEAD