FDA Adverse Event Malfunction Summary report: N

CAPSUREFIX NOVUS

MDR report key: 2131226 · Received June 16, 2011

Report

Report Number
2649622-2011-10444
Event Type
Malfunction
Date Received
June 16, 2011
Date of Event
April 25, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039/S009
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY FOR (B)(4): THE FULL LEAD WAS RETURNED, ANALYZED AND THE LEAD WAS STRETCHED. IT WAS NOTED THAT THE INNER INSULATION WAS KINKED/BUCKLED AND THE LEAD APPEARED DAMAGED AT IMPLANT. THE ANALYST ALSO COMMENTED THAT THE LEAD HAS BEEN STRETCHED, SO THE INNER INSULATION BUCKLED AND PREVENTS THE HELIX FROM EXTENDING AND RETRACTING PROPERLY.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN IMPLANT ATTEMPT, THE PHYSICIAN WAS HAVING DIFFICULTY POSITIONING THE LEAD. WHEN ATTEMPTING TO REPOSITION THE HELIX ON THE LEAD WOULD NOT RETRACT. IT WAS ALSO NOTED THAT WHEN THE LEAD WAS REMOVED FROM THE BODY AND THE STYLET REMOVED THAT THE HELIX ALSO RETRACTED. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN IMPLANT ATTEMPT, THE PHYSICIAN WAS HAVING DIFFICULTY POSITIONING THE LEAD. WHEN ATTEMPTING TO REPOSITION THE HELIX ON THE LEAD WOULD NOT RETRACT. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 5076 ASKU

Patients

Seq Age Sex Outcome Treatment
1 30 YR Other