SECURA DR
Report
- Report Number
- 6000144-2011-03453
- Event Type
- Injury
- Date Received
- June 16, 2011
- Manufacturer
- MEDTRONIC MED REL, INC.
- Product Code
- LWS
- PMA / PMN Number
- P980016/S114
- Removal / Correction Number
- ASKU
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY : (B)(4): THE ACTUAL DEVICE WAS NOT RECEIVED FOR EVALUATION, HOWEVER PERFORMANCE DATA WAS COLLECTED FROM THE DEVICE AND ANALYZED. ANALYSIS REVEALED NO ANOMALIES OBSERVED IN THE TRANSLATED SAVE TO DISK FILE, HOWEVER THE TECHNICAL SERVICES NOTE CONFIRMS T WAVE OVERSENSING AND ENGINEERING ANALYSIS CONFIRMS AN EPISODE WITH AN INSTANCE OF T WAVE OVERSENSING.
IT WAS REPORTED THAT THE PATIENT PRESENTED TO THE HOSPITAL AFTER RECEIVING SHOCKS DUE TO T-WAVE OVERSENSING. THE DEVICE WAS REPROGRAMMED AND IS STILL IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SECURA DR | IMPLANTABLE PACEMAKER/CARDIO/DEFIB | LWS | MEDTRONIC MED REL, INC. | D224DRG | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 65 YR | Hospitalization| L| R | 6935 IMPLANTABLE TACHY LEAD| 5076 IMPLANTABLE PACING LEAD| 6935 IMPLANTABLE TACHY LEAD| 5076 IMPLANTABLE PACING LEAD |