SPRINT
Report
- Report Number
- 2649622-2011-10437
- Event Type
- Injury
- Date Received
- June 16, 2011
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- LWS
- PMA / PMN Number
- P920015/S13
- Removal / Correction Number
- ASKU
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UT, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RECEIVED IN SEGMENTS. THE DISTAL CONDUCTOR APPEARED FRACTURED. BLOOD/BODY FLUID WAS FOUND ON SEVERAL CONDUCTORS (NOT OBSTRUCTED), AND BLOOD WAS FOUND IN/ON THE HELIX/LOBE MECHANISM AND THE HELIX/LOBE MECHANISM (SLEEVE HEAD). THE INNER TUBING WAS KINKED/BUCKLED, AND THE HELIX/LOBE WAS DISTORTED/BENT. THE OUTER INSULATION WAS TORN, BREACHED CUT, AND EXHIBITED COSMETIC DEPRESSION. A WHITE SUBSTANCE WAS FOUND ON THE OUTER INSULATION AND THE EXPOSED DEFIB COIL. THE LEAD APPEARED FLEXED WITHIN 5CM OF THE ANCHORING SLEEVE.
IT WAS REPORTED THAT THE SHOWED SOME SHORT V-V INTERVALS POSSIBLY DUE TO NOISE. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPRINT | IMPLANTABLE TACHY LEAD | LWS | MEDTRONIC PUERTO RICO, INC. | 6945 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 72 YR | Hospitalization| R | 7232CX IMPLANTABLE PACEMAKER/CARDIO/DEFIB |