FDA Adverse Event Injury Summary report: N

SPRINT

MDR report key: 2131181 · Received June 16, 2011

Report

Report Number
2649622-2011-10437
Event Type
Injury
Date Received
June 16, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S13
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RECEIVED IN SEGMENTS. THE DISTAL CONDUCTOR APPEARED FRACTURED. BLOOD/BODY FLUID WAS FOUND ON SEVERAL CONDUCTORS (NOT OBSTRUCTED), AND BLOOD WAS FOUND IN/ON THE HELIX/LOBE MECHANISM AND THE HELIX/LOBE MECHANISM (SLEEVE HEAD). THE INNER TUBING WAS KINKED/BUCKLED, AND THE HELIX/LOBE WAS DISTORTED/BENT. THE OUTER INSULATION WAS TORN, BREACHED CUT, AND EXHIBITED COSMETIC DEPRESSION. A WHITE SUBSTANCE WAS FOUND ON THE OUTER INSULATION AND THE EXPOSED DEFIB COIL. THE LEAD APPEARED FLEXED WITHIN 5CM OF THE ANCHORING SLEEVE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE SHOWED SOME SHORT V-V INTERVALS POSSIBLY DUE TO NOISE. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6945 ASKU

Patients

Seq Age Sex Outcome Treatment
1 72 YR Hospitalization| R 7232CX IMPLANTABLE PACEMAKER/CARDIO/DEFIB