FDA Adverse Event Injury Summary report: N

SCREW-IN

MDR report key: 2131167 · Received June 16, 2011

Report

Report Number
2649622-2011-10427
Event Type
Injury
Date Received
June 16, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
K896313
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY FOR (B)(4): ANALYSIS OF THE DEVICE REVEALED NORMAL BATTERY DEPLETION. WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA. BATTERY DEPLETION WAS INDICATED. ELECTIVE REPLACEMENT INDICATOR (ERI) LOGGED ON (B)(6) 2011.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT WAS FEELING FATIGUED WHEN THE DEVICE REACHED ELECTIVE REPLACEMENT INDICATOR. THE DEVICE WAS IMPLANTED ONLY FOUR YEARS. THE DEVICE WAS EXPLANTED AND REPLACED. IT WAS FURTHER REPORTED THAT THE RIGHT VENTRICULAR LEAD HAD HIGH THRESHOLDS. THE LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SCREW-IN IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 4058M ASKU

Patients

Seq Age Sex Outcome Treatment
1 85 YR Hospitalization| R 5068 IMPLANTABLE PACING LEAD