FDA Adverse Event
Malfunction
Summary report: N
ATTAIN STARFIX
MDR report key: 2131141
·
Received June 16, 2011
Report
- Report Number
- 2649622-2011-10407
- Event Type
- Malfunction
- Date Received
- June 16, 2011
- Date of Event
- April 30, 2011
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- DTB
- PMA / PMN Number
- P060039
- Removal / Correction Number
- ASKU
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY FOR (B)(4): THE FULL LEAD WAS RETURNED AND ANALYZED AND NO ANOMALIES WERE FOUND. THERE WAS BLOOD/BODY FLUID IN THE DISTAK CONDUCTOR AND THE OUTER TUBING OVERLAY. THERE WAS ALSO BLOOD IN/ON THE LOBE MECHANISM.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING THE IMPLANT PROCEDURE, THE LOBE OF THE LEAD COULD NOT BE OPENED, EVEN AFTER SEVERAL ATTEMPTS. THE LEAD WAS REMOVED AND REPLACED BY A NEW LEAD. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ATTAIN STARFIX | IMPLANTABLE PACING LEAD | DTB | MEDTRONIC PUERTO RICO, INC. | 4195 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 58 YR | Other |