FDA Adverse Event Injury Summary report: N

ENTRUST AT

MDR report key: 2131095 · Received June 16, 2011

Report

Report Number
6000094-2011-01133
Event Type
Injury
Date Received
June 16, 2011
Manufacturer
MEDTRONIC S.A.
Product Code
LWS
PMA / PMN Number
P980016/S46
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE. EVALUATION SUMMARY: (B)(4), THE ACTUAL DEVICE WAS NOT RECEIVED FOR EVALUATION. WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA. BATTERY VOLTAGE - PRE/APPROACHING ELECTIVE REPLACEMENT INDICATOR. WEEKLY TREND IN SAVE TO DISK DATA SHOWS MIN BAT= 2.62 VOLTS IN WEEK OF (B)(6) 2010, IS BEFORE RRT<= 2.61 VOLT.

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY FOR (B)(4): ANALYSIS REVEALED THAT THE DEVICE DID NOT MEET EXPECTED LONGEVITY, THE CAUSE IS UNKNOWN. WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA. BATTERY VOLTAGE - PRE/APPROACHING ELECTIVE REPLACEMENT INDICATOR. WEEKLY TREND IN SAVE TO DISK DATA SHOWS MIN BAT= 2.62 VOLTS IN WEEK OF (B)(6) 2010, IS BEFORE RRT<= 2.61 VOLT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE HAD PREMATURE BATTERY DEPLETION. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENTRUST AT IMPLANTABLE PACEMAKER/CARDIO/DEFIB LWS MEDTRONIC S.A. D154ATG ASKU

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| (B)(4) IMPLANTABLE PACING LEAD