FDA Adverse Event Malfunction Summary report: N

CAPSUREFIX NOVUS

MDR report key: 2131013 · Received June 16, 2011

Report

Report Number
2649622-2011-10350
Event Type
Malfunction
Date Received
June 16, 2011
Date of Event
December 13, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039/S009
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED, ANALYZED, AND THE DISTAL CONDUCTOR WAS DISTORTED. IT WAS ALSO NOTED THAT THE PROXIMAL CONDUCTOR HAD BLOOD/BODY FLUID (NOT OBSTRUCTED), THE OUTER INSULATION WAS BREACHED CUT, THERE WAS BLOOD IN/ON THE HELIX MECHANISM, AND THE LEAD WAS DAMAGED AT IMPLANT. THE ANALYST ALSO NOTED THAT THE LEAD WAS RECEIVED WITH THE HELIX FULLY RETRACTED. WHEN THE HELIX WAS RETRACTED DURING THE PROCEDURE, THE IS-1 CONNECTOR PIN WAS TURNED AN EXCESSIVE NUMBER OF TIMES RESULTING IN DISTORTION OF THE DISTAL CONDUCTOR WITHIN THE CONNECTOR.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING THE IMPLANT PROCEDURE, THERE WERE NUMEROUS ATTEMPTS TO DEPLOY AND SECURE THE LEAD IN THE RIGHT VENTRICULAR APEX. THE LEAD WAS NOT USED AND ANOTHER LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 5076 ASKU

Patients

Seq Age Sex Outcome Treatment
1 72 YR Other