SPRINT QUATTRO SECURE
Report
- Report Number
- 2649622-2011-10338
- Event Type
- Injury
- Date Received
- June 16, 2011
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- LWS
- PMA / PMN Number
- P920015/S24
- Removal / Correction Number
- Z-0475-2011
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THE DEFIBRILLATION COIL WAS DISTORTED, THE INNER TUBING WAS KINKED/BUCKLED, THERE WAS A WHITE SUBSTANCE ON THE OUTER INSULATION, THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM (SLEEVE HEAD), THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM AND THERE WAS APPARENT EXPLANT DAMAGE.
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH THRESHHOLDS AND A DECREASE IN SENSING. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPRINT QUATTRO SECURE | IMPLANTABLE TACHY LEAD | LWS | MEDTRONIC PUERTO RICO, INC. | 6947 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 34 YR | Hospitalization| R | (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB |