FDA Adverse Event Malfunction Summary report: N

ATTAIN ABILITY

MDR report key: 2130975 · Received June 16, 2011

Report

Report Number
2649622-2011-10330
Event Type
Malfunction
Date Received
June 16, 2011
Date of Event
April 25, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
OJX
PMA / PMN Number
P080006
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY : (B)(4): THE FULL LEAD WAS RETURNED AND ANALYZED; ANALYSIS REVEALED THAT THE OUTER INSULATION WAS BREACHED CUT. ALL CONDUCTORS WERE CUT AND THERE WAS BLOOD/BODY FLUID (NOT OBSTRUCTED) ON ALL CONDUCTORS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULATION OF THE LEAD WAS DAMAGED DURING INITIAL ATTEMPT TO SLIT. THE LEAD WAS NOT USED AND ANOTHER WAS SUCCESSFULLY IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ATTAIN ABILITY IMPLANTABLE PACING LEAD OJX MEDTRONIC PUERTO RICO, INC. 4196 ASKU

Patients

Seq Age Sex Outcome Treatment
1 68 YR Other 6947 IMPLANTABLE TACHY LEAD| (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 5076 IMPLANTABLE PACING LEAD