FDA Adverse Event
Malfunction
Summary report: N
ATTAIN ABILITY
MDR report key: 2130975
·
Received June 16, 2011
Report
- Report Number
- 2649622-2011-10330
- Event Type
- Malfunction
- Date Received
- June 16, 2011
- Date of Event
- April 25, 2011
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- OJX
- PMA / PMN Number
- P080006
- Removal / Correction Number
- ASKU
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY : (B)(4): THE FULL LEAD WAS RETURNED AND ANALYZED; ANALYSIS REVEALED THAT THE OUTER INSULATION WAS BREACHED CUT. ALL CONDUCTORS WERE CUT AND THERE WAS BLOOD/BODY FLUID (NOT OBSTRUCTED) ON ALL CONDUCTORS.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE INSULATION OF THE LEAD WAS DAMAGED DURING INITIAL ATTEMPT TO SLIT. THE LEAD WAS NOT USED AND ANOTHER WAS SUCCESSFULLY IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ATTAIN ABILITY | IMPLANTABLE PACING LEAD | OJX | MEDTRONIC PUERTO RICO, INC. | 4196 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 68 YR | Other | 6947 IMPLANTABLE TACHY LEAD| (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 5076 IMPLANTABLE PACING LEAD |