FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 2130971 · Received June 16, 2011

Report

Report Number
1423500-2011-07723
Event Type
Malfunction
Date Received
June 16, 2011
Date of Event
May 23, 2011
Report Date
May 23, 2011
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THIS REPORT FOR A SYSTEM ERROR 2240 (AIR IN SET) WAS CONFIRMED. THE CAUSE OF THE ALARM WAS DUE TO USER ERROR. LABEL REVIEW WAS PERFORMED AND THE REVIEW FOUND THE LABELING ADEQUATE FOR THE USER ERROR IN THE COMPLAINT. SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH RENQ-CAPA-(B)(4).

Additional Manufacturer Narrative · 1

(B)(4). DURING INVESTIGATION BY BAXTER THIS INCIDENT WAS DETERMINED TO BE CAUSED BY USE ERROR AND THERE WAS NO ALLEGATION OF A PRODUCT MALFUNCTION. THEREFORE, A BATCH REVIEW AND SAMPLE EVALUATION WILL NOT BE CONDUCTED.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BAXTER'S TECHNICAL SERVICE CENTER AND REPORTED A SYSTEM ERROR 2240 (AIR IN LINE) ALARM THAT APPEARED ON THE HOMECHOICE (HC) MACHINE DURING DWELL 4 OF 4. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) EXPLAINED ALARM AND THE HOME PATIENT (HP) SAID THAT HE HAS DISCONNECTED A FEW TIMES DURING DWELL WITHOUT PRESSING STOP. THE TSR EXPLAINED THE DISCONNECT PROCEDURE. THE TSR ASSISTED TO RETRIEVE THE CASSETTE. THE TSR ADVISED TO NOTIFY THE NURSE OF THIS EVENT. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 HOMECHOICE