FDA Adverse Event Injury Summary report: N

VIRTUOSO II DR

MDR report key: 2130958 · Received June 16, 2011

Report

Report Number
6000144-2011-03422
Event Type
Injury
Date Received
June 16, 2011
Date of Event
April 20, 2011
Manufacturer
MEDTRONIC MED REL, INC.
Product Code
LWS
PMA / PMN Number
P980016/S157
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE ACTUAL DEVICE WAS NOT RECEIVED FOR EVALUATION. WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA. A VENTRICULAR FIBRILLATION EPISODE WAS NOTED OF 170 MS ON (B)(6) 2010.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT RECEIVED INAPPROPRIATE THERAPY DUE TO RAPID ONSET OF SUPRAVENTRICULAR TACHYCARDIA. THE DEVICE WAS REPROGRAMMED AND REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VIRTUOSO II DR IMPLANTABLE PACEMAKER/CARDIO/DEFIB LWS MEDTRONIC MED REL, INC. D274DRG ASKU

Patients

Seq Age Sex Outcome Treatment
1 51 YR Life Threatening| R 7120 COMPETITOR IMPLANTABLE TACHY LEAD| 4076 IMPLANTABLE PACING LEAD