FDA Adverse Event Malfunction Summary report: N

CAPSUREFIX NOVUS

MDR report key: 2130933 · Received June 16, 2011

Report

Report Number
2649622-2011-10308
Event Type
Malfunction
Date Received
June 16, 2011
Date of Event
April 26, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039/S009
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY FOR (B)(4): THE FULL LEAD WAS RETURNED AND ANALYZED AND IT WAS FOUND THAT THERE IS INTERMITTENCY BETWEEN THE PIN AND CAP ON THE CONNECTOR. THERE WAS BLOOD IN/ON THE HELIX MECHANISM.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING THE IMPLANT PROCEDURE, THE ATRIAL LEAD HAD HIGH IMPEDANCE. THE LEAD WAS REMOVED AND REPLACED WITH ANOTHER LEAD. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 5076 ASKU

Patients

Seq Age Sex Outcome Treatment
1 71 YR Other