FDA Adverse Event Malfunction Summary report: N

CAPSUREFIX NOVUS MRI

MDR report key: 2130932 · Received June 16, 2011

Report

Report Number
2649622-2011-10307
Event Type
Malfunction
Date Received
June 16, 2011
Date of Event
April 29, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
NVN
PMA / PMN Number
P090013
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED AND ANALYZED. PRIMARY RESULTS REVEALED NO ANOMALIES FOUND. BLOOD WAS PRESENT IN/ON THE HELIX/LOBE MECHANISM AND SLEEVE HEAD.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING IMPLANT OF THE LEAD AND MULTIPLE REPOSITIONING ATTEMPTS THERE WAS HIGH IMPEDANCE DUE TO A POSSIBLE LEAD MALFUNCTION. THEREFORE THE LEAD WAS EXPLANTED AND REPLACED WITH A NEW LEAD. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS MRI IMPLANTABLE PACING LEAD NVN MEDTRONIC PUERTO RICO, INC. 5086MRI ASKU

Patients

Seq Age Sex Outcome Treatment
1 59 YR Other