CAPSUREFIX NOVUS MRI
Report
- Report Number
- 2649622-2011-10307
- Event Type
- Malfunction
- Date Received
- June 16, 2011
- Date of Event
- April 29, 2011
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- NVN
- PMA / PMN Number
- P090013
- Removal / Correction Number
- ASKU
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AL, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED AND ANALYZED. PRIMARY RESULTS REVEALED NO ANOMALIES FOUND. BLOOD WAS PRESENT IN/ON THE HELIX/LOBE MECHANISM AND SLEEVE HEAD.
IT WAS REPORTED THAT DURING IMPLANT OF THE LEAD AND MULTIPLE REPOSITIONING ATTEMPTS THERE WAS HIGH IMPEDANCE DUE TO A POSSIBLE LEAD MALFUNCTION. THEREFORE THE LEAD WAS EXPLANTED AND REPLACED WITH A NEW LEAD. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CAPSUREFIX NOVUS MRI | IMPLANTABLE PACING LEAD | NVN | MEDTRONIC PUERTO RICO, INC. | 5086MRI | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 59 YR | Other |