SPRINT QUATTRO SECURE
Report
- Report Number
- 2649622-2011-10286
- Event Type
- Malfunction
- Date Received
- June 16, 2011
- Date of Event
- April 22, 2011
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- LWS
- PMA / PMN Number
- P920015/S24
- Removal / Correction Number
- Z-0475-2011
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TU
- Reporter Occupation
- PHYSICIAN
Narratives
THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY FOR (B)(4): THE FULL LEAD WAS RETURNED, ANALYZED AND THE HELIX WAS DISTORTED/BENT. IT WAS NOTED THAT THERE WAS BLOOD IN/ON THE HELIX MECHANISM (SLEEVE HEAD) AND THE LEAD APPEARED DAMAGED AT IMPLANT. THE ANALYST COMMENTED THAT THE HELIX WILL NOT EXTEND AND RETRACT AT THIS TIME DUE TO THE LEAD BEING RETURNED WITH DAMAGED HELIX.
IT WAS REPORTED THAT DURING IMPLANT, IT WAS OBSERVED THAT THE RIGHT VENTRICULAR LEAD HELIX GOT STUCK. THE HELIX SEEMED TO BE DEFORMED. THE LEAD WAS REMOVED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPRINT QUATTRO SECURE | IMPLANTABLE TACHY LEAD | LWS | MEDTRONIC PUERTO RICO, INC. | 6947 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 63 YR | Other |