FDA Adverse Event Malfunction Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 2130884 · Received June 16, 2011

Report

Report Number
2649622-2011-10286
Event Type
Malfunction
Date Received
June 16, 2011
Date of Event
April 22, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S24
Removal / Correction Number
Z-0475-2011
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TU
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY FOR (B)(4): THE FULL LEAD WAS RETURNED, ANALYZED AND THE HELIX WAS DISTORTED/BENT. IT WAS NOTED THAT THERE WAS BLOOD IN/ON THE HELIX MECHANISM (SLEEVE HEAD) AND THE LEAD APPEARED DAMAGED AT IMPLANT. THE ANALYST COMMENTED THAT THE HELIX WILL NOT EXTEND AND RETRACT AT THIS TIME DUE TO THE LEAD BEING RETURNED WITH DAMAGED HELIX.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING IMPLANT, IT WAS OBSERVED THAT THE RIGHT VENTRICULAR LEAD HELIX GOT STUCK. THE HELIX SEEMED TO BE DEFORMED. THE LEAD WAS REMOVED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6947 ASKU

Patients

Seq Age Sex Outcome Treatment
1 63 YR Other