FDA Adverse Event Injury Summary report: N

ENRHYTHM DR

MDR report key: 2130827 · Received June 16, 2011

Report

Report Number
6000094-2011-01098
Event Type
Injury
Date Received
June 16, 2011
Manufacturer
MEDTRONIC S.A.
Product Code
DXY
PMA / PMN Number
P980035/S38
Removal / Correction Number
Z-1440-2010
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY : (B)(4): THE ACTUAL DEVICE WAS NOT RECEIVED FOR EVALUATION, HOWEVER PERFORMANCE DATA WAS COLLECTED FROM THE DEVICE AND ANALYZED. ANALYSIS REVEALED BATTERY DEPLETION WAS INDICATED (ERI) AND ERI WAS DUE TO HIGH BATTERY IMPEDANCE LOGGED ON (B)(6) 2010, PRIOR TO EXPLANT. RAMWARE VERSION 8 INSTALLED ON (B)(6) 2010. HIGH BATTERY IMPEDANCE LED TO ERI.

Description of Event or Problem · 1

IT WAS REPORTED THAT DEVICE REACHED ELECTIVE REPLACEMENT INDICATOR (ERI) AND THE DEVICE HAD PREMATURE DEPLETION. THE DEVICE WAS PLANNED TO BE REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENRHYTHM DR IMPLANTABLE PULSE GENERATOR DXY MEDTRONIC S.A. P1501DR ASKU

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R