PERCLOSE¿ PROSTYLE¿
Report
- Report Number
- 2024168-2025-01303
- Event Type
- Injury
- Date Received
- February 4, 2025
- Date of Event
- November 1, 2024
- Report Date
- April 1, 2025
- Manufacturer
- ABBOTT VASCULAR INC.
- Product Code
- MGB
- UDI-DI
- 08717648344039
- PMA / PMN Number
- P960043
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
ANALYSIS WAS PERFORMED ON THE RETURNED DEVICE. THE REPORTED NEEDLE TO CUFF MISS WAS CONFIRMED. PRODUCTION RECORD AND CORRECTIVE AND PREVENTIVE ACTIONS (CAPA) REVIEWS WERE PERFORMED AND REVEALED NO INDICATION OF A PRODUCT QUALITY ISSUE. THE REPORTED DIFFICULTY AND SUBSEQUENT TREATMENT APPEAR TO BE RELATED TO AN INTERACTION OF THE DEVICE WITH PATIENT ANATOMY OR INABILITY TO MAINTAIN POSITION/STABILITY OF THE DEVICE DURING DEPLOYMENT DUE TO CIRCUMSTANCES OF THE PROCEDURE. THERE IS NO INDICATION OF A PRODUCT QUALITY ISSUE WITH RESPECT TO THE DESIGN, MANUFACTURE, OR LABELING OF THE DEVICE. CORRECTION: D4 - LOT #: UPDATED FROM 3071141 TO 3070241. D4 - PRIMARY UDI NUMBER: UPDATED FROM (B)(4). H4 - DEVICE MFG DATE: UPDATED FROM 7/11/2023 TO 7/2/2023. H6- MEDICAL DEVICE PROBLEM CODE: CODE 2616 WAS REMOVED AND CODE 1142 WAS ADDED.
MANUFACTURER'S INVESTIGATION IS STILL PENDING AT THIS TIME. RESULTS AND CONCLUSIONS WILL BE PROVIDED IN THE FINAL REPORT.
IT WAS REPORTED THAT THIS WAS AN ARTERIOTOMY CLOSURE OF A RIGHT COMMON FEMORAL ARTERY USING A PROSTYLE RELATED TO AN UNKNOWN PROCEDURE. REPORTEDLY, AFTER REMOVING THE PLUNGER THE SUTURE CAME OUT OF THE NEEDLE AND THE KNOT WAS MADE OUTSIDE OF THE PATIENT. AN UNKNOWN METHOD WAS USED TO ACHIEVE HEMOSTASIS. THERE WAS NO ADVERSE PATIENT SEQUELA AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 61281 | PERCLOSE¿ PROSTYLE¿ | DEVICE, HEMOSTASIS, VASCULAR | MGB | ABBOTT VASCULAR INC. | 12773-02 | 3070241 | 08717648344039 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | Required Intervention |