FDA Adverse Event
Injury
Summary report: N
CAPSUREFIX NOVUS MRI
MDR report key: 2130806
·
Received June 16, 2011
Report
- Report Number
- 2649622-2011-10256
- Event Type
- Injury
- Date Received
- June 16, 2011
- Date of Event
- April 6, 2011
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- NVN
- PMA / PMN Number
- P090013
- Removal / Correction Number
- ASKU
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FS
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED AND ANALYSIS REVEALED THAT THE HELIX/LOBE WAS DISTORTED/BENT AND THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM. VISUAL ANALYSIS REVEALED THAT A GUIDE TOOTH WAS DAMAGED AND THAT THE LEAD WAS DAMAGED AT IMPLANT.
Description of Event or Problem · 1
IT WAS REPORTED THAT THERE WAS DISLODGEMENT OF THE ATRIAL LEAD DUE TO THE QUICK EXTENSION OF THE HELIX. THE LEAD WAS REMOVED AND ANOTHER WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CAPSUREFIX NOVUS MRI | IMPLANTABLE PACING LEAD | NVN | MEDTRONIC PUERTO RICO, INC. | 5086MRI | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization| R |