FDA Adverse Event Injury Summary report: N

CD HORIZON SPINAL SYSTEM

MDR report key: 2130743 · Received June 16, 2011

Report

Report Number
1030489-2011-00741
Event Type
Injury
Date Received
June 16, 2011
Report Date
May 17, 2011
Manufacturer
MEDTRONIC SOFAMOR DANEK
Product Code
KWP
PMA / PMN Number
K052187
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED FOR EVALUATION. VISUAL AND OPTICAL EXAMINATION OF RETURNED SET SCREW ROD INTERFACE SURFACES EXHIBIT SECONDARY WEAR / DAMAGE WHICH INHIBIT DETERMINATION OF ROOT CAUSE WITH RESPECT TO ROD/SET SCREW INTERFACE LOOSENING. WITNESS MARKS NOTED ON MAS CROWN TYPICAL OF FULLY TIGHTENED SET SCREW. DAMAGE NOTED ON EXTERNAL SURFACES OF MULTI-AXIAL SCREWS. THE NATURE AND LOCATION OF THE MAS SURFACE DAMAGE IS CONSISTENT WITH THE CORRESPONDING DAMAGE NOTED ON SET SCREW ROD INTERFACE SURFACES. WITNESS MARKS NOTED ON MAS CROWN TYPICAL OF FULLY TIGHTENED SET SCREW. SET SCREW INTERFACE WITNESS MARKS AT L4 (LOCATION OF SET SCREW DISENGAGEMENT) VISUALLY APPEAR TO BE LESS PRONOUNCED WHEN COMPARED TO THE L5 AND S1 AND THE SAMPLE COCR PLUS ROD SET SCREW ROD INTERFACE WITNESS MARKS. ADDITIONAL ABRASIONS ALSO NOTED CONSISTENT WITH ROD MOVEMENT AFTER SET SCREW LOOSENED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT UNDERWENT A TRANSFORAMINAL LUMBAR INTERBODY FUSION AT L4-S1. IT WAS REPORTED THAT THE PATIENT HEARD A POP SOMETIME POST-OP. DURING A ROUTINE FOLLOW UP VISIT 6 WEEKS POST-OP IT WAS FOUND THAT THE SET SCREWS HAD BACKED OUT OF THE BONE SCREWS AT L4. THE RODS MOVED POSTERIORLY AND CAUDAL. THE PATIENT UNDERWENT A REVISION SURGERY WITH THE REPLACEMENT OF THE RODS, L4 BONE SCREWS, CROSSLINK AND SET SCREWS. NO ADDITIONAL PATIENT COMPLICATIONS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CD HORIZON SPINAL SYSTEM APPLIANCE, FIXATION, SPINAL INTERLAMINAL KWP MEDTRONIC SOFAMOR DANEK NA NI

Patients

Seq Age Sex Outcome Treatment
1 00040 YR Required Intervention ROD, SCREWS, CROSSLINK