FDA Adverse Event
Malfunction
Summary report: N
MAXIMO VR
MDR report key: 2130690
·
Received June 16, 2011
Report
- Report Number
- 6000144-2011-03371
- Event Type
- Malfunction
- Date Received
- June 16, 2011
- Manufacturer
- MEDTRONIC MED REL, INC.
- Product Code
- LWS
- PMA / PMN Number
- P980016/S37
- Removal / Correction Number
- ASKU
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- ID, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY (B)(4) THE ACTUAL DEVICE WAS NOT RECEIVED FOR EVALUATION. WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA. THE DATA SHOWED A POWER ON RESET (POR) PARAMETERS FOR WRITE TO LOCKED RAM ON (B)(6) 2011 19:10:11. PATIENT ALERT TRIGGERED FOR POR ON (B)(6) 2011 18:10:11.
Description of Event or Problem · 1
IT WAS REPORTED THAT "THE DEVICE HAS BEEN ALARMING FOR ABOUT A MONTH" AND THAT THERE WAS A POWER ON RESET OF THE DEVICE. THE DEVICE REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MAXIMO VR | IMPLANTABLE PACEMAKER/CARDIO/DEFIB | LWS | MEDTRONIC MED REL, INC. | 7232CX | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 63 YR | Other | (B)(4) IMPLANTABLE TACHY LEAD |