FDA Adverse Event Injury Summary report: N

INSPIRA SMOOTH SILICONE GEL FILLED BREAST IMPLANT

MDR report key: 21305576 · Received February 4, 2025

Report

Report Number
9617229-2025-01901
Event Type
Injury
Date Received
February 4, 2025
Date of Event
November 14, 2024
Report Date
April 10, 2025
Manufacturer
ALLERGAN (COSTA RICA)
Product Code
FTR
PMA / PMN Number
P020056
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
OH
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL, CHANGED, AND/OR CORRECTED DATA: A2 B5 D6B D9 H3 H6. LABORATORY ANALYSIS SUMMARY- THE DEVICE RELATED TO THE REPORTED EVENTS CAPSULAR CONTRACTURE WAS RECEIVED ON MARCH 18, 2025 WITH LOT NUMBER 1210016. BASED ON THE PRODUCT ANALYSIS PERFORMED, THE ASSESSMENTS OF THE COMPLAINT CODES ARE: -CAPSULAR CONTRACTURE: UNABLE TO OBSERVE. AS PER THE INVESTIGATION PROCEDURE, CREASES AND WEAR ABRASION WERE OBSERVED AND NONE OF THE OBSERVATIONS ARE FOUND TO BE POTENTIALLY RELATED TO THE MANUFACTURING PROCESS, NO FURTHER ACTIONS ARE REQUIRED.

Additional Manufacturer Narrative · 0

THE EVENT OF CAPSULAR CONTRACTURE IS A PHYSIOLOGICAL COMPLICATION AND ANALYSIS OF THE DEVICE GENERALLY DOES NOT ASSIST ALLERGAN IN DETERMINING A PROBABLE CAUSE FOR THIS EVENT. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. REASON FOR REOPERATION: CAPSULAR CONTRACTURE BAKER GRADE IV. A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED.

Description of Event or Problem · 0

THE DEVICE HAS BEEN EXPLANTED.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL REPORTED "CAPSULAR CONTRACTURES BAKER GRADE IV" THIS IS FOR THE RIGHT-SIDE DEVICE. THE DEVICE REMAINS IMPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
65167 INSPIRA SMOOTH SILICONE GEL FILLED BREAST IMPLANT PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED FTR ALLERGAN (COSTA RICA) 1210016

Patients

Seq Age Sex Outcome Treatment
1 51 YR Female Required Intervention