BD PHOENIX¿ NMIC-306
Report
- Report Number
- 1119779-2025-00030
- Event Type
- Malfunction
- Date Received
- February 4, 2025
- Date of Event
- December 17, 2024
- Report Date
- April 3, 2025
- Manufacturer
- BECTON DICKINSON & CO. (SPARKS)
- Product Code
- LON
- UDI-DI
- 30382904492923
- PMA / PMN Number
- SEE H11
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THERE WERE MULTIPLE 510K NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR THE ADDITIONAL 510K IS AS FOLLOWS: G4. PMA / 510(K)#: K020322, K022129, K023444, K023634, K023858, K024153, K031530, K031699, K031912, K032299, K032567, K032655, K033362, K033560, K041384, K042932, K052269, K060214, K060217, K060257, K060444, K060447, K061327, K061355, K062207, K062944, K063301, K063486, K063573, K063811, K063824, K071623, K132674, K132909, K151320, K181665, K173252, K190905, K163637, K173523, K033458, AND K123266. H3. A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
INVESTIGATION SUMMARY - THIS COMPLAINT IS FOR HIGH MIC MEROPENEM (MEM) WITH KLEBSIELLA PNEUMONIAE, ESCHERICHIA COLI AND ENTEROBACTER CLOACAE COMPLEX WHEN USING PHOENIX PANEL NMIC-306 (CATALOG NUMBER 449292) BATCH NUMBER 4317031. THE CUSTOMER DID NOT RETURN PANELS BUT RETURNED ISOLATES, PHOENIX GENERATED LAB REPORTS AND BINARY FILES FOR THE INVESTIGATION. THE PHOENIX GENERATED LAB REPORTS SHOW HIGH MIC MEM WITH KLEBSIELLA PNEUMONIAE, ESCHERICHIA COLI AND ENTEROBACTER CLOACAE COMPLEX WHEN USING THE COMPLAINT BATCH. TO INVESTIGATE, RETENTION PANELS OF THE COMPLAINT BATCH WERE TESTED USING CUSTOMER RETURNED ISOLATES KLEBSIELLA PNEUMONIAE N820901067, E. COLI N825007857, E. CLOACAE COMPLEX N824019728, E. CLOACAE COMPLEX N823011694, K. PNEUMONIAE N824011063, K. PNEUMONIAE N823011300 AND E. COLI N825015335 ON A PHOENIX M50 MACHINE AND EVALUATED FOR MEM MIC RESULTS. ALSO, CONTROL PANELS FROM THE SAME MATERIAL BUT DIFFERENT BATCH WERE TESTED USING CUSTOMER RETURNED ISOLATES KLEBSIELLA PNEUMONIAE N820901067, E. COLI N825007857, E. CLOACAE COMPLEX N824019728, E. CLOACAE COMPLEX N823011694, K. PNEUMONIAE N824011063, K. PNEUMONIAE N823011300 AND E. COLI N825015335 ON A PHOENIX M50 MACHINE AND EVALUATED FOR MEM MIC RESULTS. ALL PANELS TESTED RETURNED THE EXPECTED SENSITIVE MICS FOR MEM, THIS COMPLAINT IS NOT CONFIRMED. A REVIEW OF THE BINARY FILES WAS DETERMINED NOT TO BE REQUIRED BASED ON THE RESULTS OF THE INVESTIGATION. THE BATCH HISTORY RECORD WAS SATISFACTORY AND NO QUALITY NOTIFICATIONS WERE GENERATED DURING MANUFACTURING AND INSPECTION. COMPLAINT TRENDING WAS PERFORMED AND NO TRENDS WERE IDENTIFIED ASSOCIATED WITH THIS DEFECT. BD WILL CONTINUE TO MONITOR FOR TRENDS AND TAKE ACTION AS NECESSARY. PLEASE CONTINUE TO COMMUNICATE ANY ADDITIONAL CONCERNS.
IT WAS REPORTED WHILE USING THE BD PHOENIX¿ NMIC-306 AN UNSPECIFIED NUMBER OF PATIENT ISOLATES HAD HIGH MICS FOR THE DRUG MEROPENEM. CONFIRMATORY TESTING WAS PERFORMED USING E-TEST STRIP. NO HEALTH IMPACT OR CONSEQUENCE REPORTED.
IT WAS REPORTED WHILE USING THE BD PHOENIX¿ NMIC-306 AN UNSPECIFIED NUMBER OF PATIENT ISOLATES HAD HIGH MICS FOR THE DRUG MEROPENEM. CONFIRMATORY TESTING WAS PERFORMED USING E-TEST STRIP. NO HEALTH IMPACT OR CONSEQUENCE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 64153 | BD PHOENIX¿ NMIC-306 | SYSTEM, TEST, AUTOMATED ANTIMICROBIAL SUSCEPTIBILITY, SHORT INCUBATION | LON | BECTON DICKINSON & CO. (SPARKS) | 4317031 | 30382904492923 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |