ATTAIN ABILITY
Report
- Report Number
- 2649622-2011-10172
- Event Type
- Injury
- Date Received
- June 16, 2011
- Date of Event
- February 18, 2011
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- OJX
- PMA / PMN Number
- P080006
- Removal / Correction Number
- ASKU
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT THE LEFT VENTRICULAR LEAD CAUSED EXTRACARDIAC STIMULATION, SO THE LEAD WAS PROGRAMMED OFF. THE LEAD WAS TURNED BACK ON THE POLARITY WAS CHANGED WHICH RESOLVED THE EXTRACARDIAC STIMULATION AND RESULTED IN AN ELEVATED THRESHOLD. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE LEAD DISLODGED AND WAS SUBSEQUENTLY REPOSITIONED.
IT WAS REPORTED THAT THE LEFT VENTRICULAR LEAD CAUSED EXTRACARDIAC STIMULATION, SO THE LEAD WAS PROGRAMMED OFF. THE LEAD WAS TURNED BACK ON THE POLARITY WAS CHANGED WHICH RESOLVED THE EXTRACARDIAC STIMULATION AND RESULTED IN AN ELEVATED THRESHOLD. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ATTAIN ABILITY | IMPLANTABLE PACING LEAD | OJX | MEDTRONIC PUERTO RICO, INC. | 4196 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 65 YR | Hospitalization| R | 1591 COMPETITOR IMPLANTABLE TACHY LEAD| 1591 COMPETITOR IMPLANTABLE TACHY LEAD| 1688TC COMPETITOR IMPLANTABLE PACING LEAD| (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 1688TC COMPETITOR IMPLANTABLE PACING LEAD| (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB |