FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2130474 · Received June 16, 2011

Report

Report Number
2531779-2011-04232
Event Type
Malfunction
Date Received
June 16, 2011
Date of Event
April 29, 2011
Report Date
April 29, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CORRECTION NUMBER: 2531779-03/24/2010-003-R. THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2011 WITH THE FOLLOWING FINDINGS: DURING EVALUATION THE FORCE SENSOR WAS FOUND TO BE OUT OF CALIBRATION. ADDITIONALLY CONTAMINATION WAS OBSERVED ON THE FORCE SENSOR ASSEMBLY. A REVIEW OF THE PUMP HISTORY SHOWS LOSS OF PRIME WARNINGS. DURING EVALUATION THE PUMP WAS ABLE TO REWIND, LOAD AND PRIME WITH NO ISSUES. THE PUMP WAS EXERCISED FOR 24 HOURS WITH NO ISSUES.

Description of Event or Problem · 1

THE PUMP WAS RETURNED TO ANIMAS DUE TO MULTIPLE LOSS OF PRIMES. DURING EVALUATION THE FORCE SENSOR WAS FOUND TO BE OUT OF CALIBRATION. ADDITIONALLY CONTAMINATION WAS FOUND ON THE FORCE SENSOR. THIS IS BEING REPORTED BASED ON INVESTIGATION RESULTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 61 YR