FDA Adverse Event Malfunction Summary report: N

MICRO SAGITTAL SAW

MDR report key: 2130389 · Received May 9, 2011

Report

Report Number
1811755-2011-01659
Event Type
Malfunction
Date Received
May 9, 2011
Date of Event
April 11, 2011
Report Date
April 12, 2011
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
HBE
PMA / PMN Number
K032303
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IA, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED TO THE MFR FOR EVAL. THE COMPLAINT WAS DUPLICATED, THE HANDPIECE HEATED UP DUE TO CORROSION. THE DEVICE WAS REPAIRED AND RETURNED TO THE CUSTOMER.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE ITEM GOT HOT AND THEN STOPPED. NO OTHER INFO WAS GIVEN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MICRO SAGITTAL SAW HBE STRYKER INSTRUMENTS KALAMAZOO 0403600583

Patients

Seq Age Sex Outcome Treatment
1 UNK