CORE IMPACTION DRILL

MDR report key: 2130365 · Received May 9, 2011

Report

Report Number
1811755-2011-01662
Event Type
Malfunction
Date Received
May 9, 2011
Date of Event
April 15, 2011
Report Date
April 15, 2011
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
ERL
PMA / PMN Number
K040300
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE HANDPIECE WAS RETURNED TO THE MFR FOR EVAL. DURING TESTING IT WAS IDENTIFIED THAT THE HANDPIECE EXCEEDED THE MAXIMUM ALLOWED TEMPERATURE RISE. THE CONTACT PINS WERE BURNT, AND THERE WAS INSUFFICIENT LUBE IN THE BEARING ON THE ROTOR AND DRIVESHAFT. THE DEVICE WAS REPAIRED AND RETURNED TO THE CUSTOMER.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PRODUCT WAS RUNNING HOT. NO OTHER INFO WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CORE IMPACTION DRILL ERL STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK