FDA Adverse Event
Malfunction
Summary report: N
CORE IMPACTION DRILL
MDR report key: 2130365
·
Received May 9, 2011
Report
- Report Number
- 1811755-2011-01662
- Event Type
- Malfunction
- Date Received
- May 9, 2011
- Date of Event
- April 15, 2011
- Report Date
- April 15, 2011
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- ERL
- PMA / PMN Number
- K040300
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE HANDPIECE WAS RETURNED TO THE MFR FOR EVAL. DURING TESTING IT WAS IDENTIFIED THAT THE HANDPIECE EXCEEDED THE MAXIMUM ALLOWED TEMPERATURE RISE. THE CONTACT PINS WERE BURNT, AND THERE WAS INSUFFICIENT LUBE IN THE BEARING ON THE ROTOR AND DRIVESHAFT. THE DEVICE WAS REPAIRED AND RETURNED TO THE CUSTOMER.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE PRODUCT WAS RUNNING HOT. NO OTHER INFO WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CORE IMPACTION DRILL | ERL | STRYKER INSTRUMENTS KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |