FDA Adverse Event Injury Summary report: N

CAPSUREFIX NOVUS

MDR report key: 2130344 · Received June 16, 2011

Report

Report Number
2649622-2011-10150
Event Type
Injury
Date Received
June 16, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039/S009
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT HAD INCREASING HIGH IMPEDANCES ON THE RIGHT VENTRICULAR LEAD FOR SEVERAL MONTHS. THE THRESHOLD CAPTURE WAS ALSO BEGINNING TO INCREASE. THE LEAD WAS CAPPED AND REPLACED. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 5076 ASKU

Patients

Seq Age Sex Outcome Treatment
1 76 YR Hospitalization| R E2DR01 IMPLANTABLE PULSE GENERATOR| 5076 X2 IMPLANTABLE PACING LEAD