STELLANT DUAL CT INJECTOR
Report
- Report Number
- 2520313-2025-00003
- Event Type
- Injury
- Date Received
- February 4, 2025
- Date of Event
- January 7, 2025
- Report Date
- March 4, 2025
- Manufacturer
- BAYER MEDICAL CARE INC
- Product Code
- DXT
- PMA / PMN Number
- K082905
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
A SYSTEM SERVICE CHECK OF THE MEDRAD® STELLANT CT INJECTION SYSTEM, SERIAL NUMBER: (B)(6), WAS PERFORMED ON (B)(6) 2025, WHICH CONFIRMED THAT THE INJECTOR WAS OPERATING WITHIN BAYER SPECIFICATIONS. THE DISPOSABLES THAT WERE IN USE DURING THE PROCEDURE WERE DISCARDED BY THE SITE; THEREFORE, THEY ARE NOT AVAILABLE FOR EVALUATION. THE CUSTOMER WAS UNABLE TO PROVIDE THE LOT NUMBER OF THE DISPOSABLES USED DURING THE PROCEDURE; THEREFORE, TESTING OF RETAINED SAMPLES WAS NOT POSSIBLE. THE OFFER OF ADDITIONAL CLINICAL APPLICATIONS TRAINING WAS DECLINED BY THE CUSTOMER. BAYER MEDICAL CARE RECEIVED INFORMATION FROM THE CUSTOMER CONFIRMING THEY BELIEVE OPERATOR ERROR IS LIKELY AND THAT THIS AIR INJECTION OCCURRED ON THE SAME DAY AND DURING THE SAME STAFF SHIFT AS ANOTHER PRIORLY REPORTED AIR INJECTION, ON (B)(6) 2025 (MFR REPORT NUMBER: 2520313-2025-00004). THE MEDRAD® STELLANT CT INJECTION SYSTEM OPERATION MANUAL CAUTIONS THE USER AS FOLLOWS: WARNING: AIR EMBOLISM HAZARD - SERIOUS PATIENT INJURY OR DEATH MAY RESULT. ENSURE PATIENT IS NOT CONNECTED WHILE PURGING AIR FROM SYRINGE, OR ENGAGING OR ADVANCING PLUNGER. EXPEL ALL TRAPPED AIR FROM THE SYRINGE(S), CONNECTORS, TUBING, AND CATHETER BEFORE CONNECTING THE SYSTEM TO THE PATIENT. TO MINIMIZE AIR EMBOLIZATION RISKS, ENSURE THAT ONE OPERATOR IS DESIGNATED THE RESPONSIBILITY OF FILLING THE SYRINGE(S). DO NOT CHANGE OPERATORS DURING THE PROCEDURE. IF AN OPERATOR CHANGE MUST OCCUR, ENSURE THAT THE NEW OPERATOR VERIFIES THAT THE FLUID PATH IS PURGED OF AIR. THIS INFORMATION DOES NOT CONSTITUTE AN ADMISSION THAT THE DEVICE, THE COMPANY, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO A REPORTABLE EVENT. NOTE: THE MANUFACTURE DATE FOR THIS DEVICE IS SEPTEMBER 29, 2005 WHICH WAS PRIOR TO THE UDI IMPLEMENTATION DATE OF SEPTEMBER 24, 2016 FOR CLASS II DEVICES. AS SUCH, THIS EQUIPMENT WAS NOT YET REQUIRED TO BEAR UDI MARKING AND/OR REPORTED INTO THE GUDID DATABASE AT THE TIME.
A SYSTEM SERVICE CHECK OF THE MEDRAD® STELLANT CT INJECTION SYSTEM, SERIAL NUMBER (B)(6), WAS PERFORMED ON (B)(6), 2025, WHICH CONFIRMED THAT THE INJECTOR WAS OPERATING WITHIN BAYER SPECIFICATIONS. THE DISPOSABLES THAT WERE IN USE DURING THE PROCEDURE WERE DISCARDED BY THE SITE; THEREFORE, THEY ARE NOT AVAILABLE FOR EVALUATION. THE CUSTOMER WAS UNABLE TO PROVIDE THE LOT NUMBER OF THE DISPOSABLES USED DURING THE PROCEDURE; THEREFORE, TESTING OF RETAINED SAMPLES WAS NOT POSSIBLE. THE OFFER OF ADDITIONAL CLINICAL APPLICATIONS TRAINING WAS DECLINED BY THE CUSTOMER. WE HAVE REACHED OUT TO THE CUSTOMER FOR ADDITIONAL INFORMATION. THIS INVESTIGATION IS ONGOING. THE MEDRAD® STELLANT CT INJECTION SYSTEM OPERATION MANUAL CAUTIONS THE USER AS FOLLOWS: WARNING: AIR EMBOLISM HAZARD - SERIOUS PATIENT INJURY OR DEATH MAY RESULT. ENSURE PATIENT IS NOT CONNECTED WHILE PURGING AIR FROM SYRINGE OR ENGAGING OR ADVANCING PLUNGER. EXPEL ALL TRAPPED AIR FROM THE SYRINGE(S), CONNECTORS, TUBING, AND CATHETER BEFORE CONNECTING THE SYSTEM TO THE PATIENT. TO MINIMIZE AIR EMBOLIZATION RISKS, ENSURE THAT ONE OPERATOR IS DESIGNATED THE RESPONSIBILITY OF FILLING THE SYRINGE(S). DO NOT CHANGE OPERATORS DURING THE PROCEDURE. IF AN OPERATOR CHANGE MUST OCCUR, ENSURE THAT THE NEW OPERATOR VERIFIES THAT THE FLUID PATH IS PURGED OF AIR. THIS INFORMATION DOES NOT CONSTITUTE AN ADMISSION THAT THE DEVICE, THE COMPANY, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO A REPORTABLE EVENT.
BAYER MEDICAL CARE INC. WAS INFORMED THAT A 69-YEAR-OLD MALE UNDERGOING A CT OF THE CHEST, ABDOMEN, AND PELVIS EXPERIENCED AN ALLEGED AIR INJECTION WHILE CONNECTED TO A MEDRAD® STELLANT CT INJECTION SYSTEM (SN: (B)(6). FOLLOWING THE PROCEDURE, THE CUSTOMER REPORTED THAT AIR WAS VIEWED ON THE DISPLAYED IMAGES WITHIN THE LEFT BRACHIOCEPHALIC VEIN, THE RIGHT ATRIUM, AND THE RIGHT VENTRICLE. THE PATIENT WAS SENT FOR A REPEAT NON-CONTRAST CT OF THE CHEST. THE CUSTOMER HAD REPORTED THAT THE AIR HAD RESOLVED, THE PATIENT IS ASYMPTOMATIC AND IS DOING WELL.
BAYER MEDICAL CARE INC. WAS INFORMED THAT A 69-YEAR-OLD MALE UNDERGOING A CT OF THE CHEST, ABDOMEN, AND PELVIS EXPERIENCED AN ALLEGED AIR INJECTION WHILE CONNECTED TO A MEDRAD® STELLANT CT INJECTION SYSTEM (SN (B)(6)). FOLLOWING THE PROCEDURE, THE CUSTOMER REPORTED THAT AIR WAS VIEWED ON THE DISPLAYED IMAGES WITHIN THE LEFT BRACHIOCEPHALIC VEIN, THE RIGHT ATRIUM, AND THE RIGHT VENTRICLE. THE PATIENT WAS SENT FOR A REPEAT NON-CONTRAST CT OF THE CHEST. THE CUSTOMER HAD REPORTED THAT THE AIR HAD RESOLVED, THE PATIENT IS ASYMPTOMATIC AND IS DOING WELL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1245827 | STELLANT DUAL CT INJECTOR | CT INJECTION SYSTEM | DXT | BAYER MEDICAL CARE INC | 59879956 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 69 YR | Male | Other |