FDA Adverse Event
Malfunction
Summary report: N
I-STAT TROPONIN (CTNI)
MDR report key: 2130287
·
Received May 9, 2011
Report
- Report Number
- 2245578-2011-00066
- Event Type
- Malfunction
- Date Received
- May 9, 2011
- Date of Event
- March 21, 2011
- Report Date
- May 6, 2011
- Manufacturer
- ABBOTT POINT OF CARE CANADA LTD.
- Product Code
- MMI
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
(B)(4). CUSTOMER RETURNED CARTRIDGES WERE VISUALLY INSPECTED AND CONFIRMED CUSTOMER COMPLAINT. TAPE GASKET HOLE CUT FOR THE SAMPLE ENTRY WELL WAS NOT PRESENT WHICH PREVENTED SAMPLE FROM ENTERING THE CARTRIDGE.
Description of Event or Problem · 1
ON (B)(6) 2011, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER WHO REPORTED THAT I-STAT TROPONIN CARTRIDGES WERE DIFFICULT TO FILL DUE TO A PLASTIC PIECE COVERING THE SAMPLE WELL. BASED ON THE INFORMATION AT THE TIME, THERE WAS NO INJURY ASSOCIATED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | I-STAT TROPONIN (CTNI) | CTNI CARTRIDGE | MMI | ABBOTT POINT OF CARE CANADA LTD. | T11016A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |